eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter User Manual

Page of 56
Instruction Manual 
 
CON 400/410 
- 48 - 
16  ADDENDUM 4: METER FACTORY DEFAULT 
S E T T I N GS  
Type Parameter 
Default 
Remarks 
P1.0 
Memory clear 
No 
Retains current memory 
P2.1 
Viewing previous 
–– 
No calibration data for 1
st
 range 
P2.2 
calibration data 
–– 
No calibration data for 2
nd
 range 
P2.3 
 
–– 
No calibration data for 3
rd
 range 
P2.4 
 
–– 
No calibration data for 4
th
 range 
P2.5 
 
–– 
No calibration data for 5
th
 range 
P3.1 
Viewing probe data 
1.0 
No offset for effective cell constant (1
st
 range) 
P3.2 
 
1.0 
No offset for effective cell constant (2
nd
 range) 
P3.3 
 
1.0 
No offset for effective cell constant (3
rd
 range) 
P3.4 
 
1.0 
No offset for effective cell constant (4
th
 range) 
P3.5 
 
1.0 
No offset for effective cell constant (5
th
 range) 
P4.1 
Ready indicator / auto 
endpoint 
READY/ON Ready 
indicator on; auto endpoint off 
P4.2 Select 
°C/°F 
°C 
Degrees 
Celsius 
P4.3 
ATC on or off 
ATC on 
–– 
P4.4 
TDS factor 
0.50 
Adjustable from 0.40 to 1.00 (available in TDS 
400 and CON 410 meters only
P5.1 
Temperature coefficient 
2.1 % per ° 
Adjustable from 0.0 to 10% 
P5.2 
Normalisation temperature 
25° C 
Adjustable from 15 to 30° C 
P6.1 
Cell constant 
1.0 
Select from k=1.0, 0.1 or 10 
P8.0 
Factory default 
No 
Retains your current settings 
 
NOTES: 
The P7.0 clock function retains the year and time that you set, even after reverting to factory 
default settings.