eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter Manuale Utente
Instruction Manual
CON 400/410
- 48 -
16 ADDENDUM 4: METER FACTORY DEFAULT
S E T T I N GS
Type Parameter
Default
Remarks
P1.0
Memory clear
No
Retains current memory
P2.1
Viewing previous
––
No calibration data for 1
st
range
P2.2
calibration data
––
No calibration data for 2
nd
range
P2.3
––
No calibration data for 3
rd
range
P2.4
––
No calibration data for 4
th
range
P2.5
––
No calibration data for 5
th
range
P3.1
Viewing probe data
1.0
No offset for effective cell constant (1
st
range)
P3.2
1.0
No offset for effective cell constant (2
nd
range)
P3.3
1.0
No offset for effective cell constant (3
rd
range)
P3.4
1.0
No offset for effective cell constant (4
th
range)
P3.5
1.0
No offset for effective cell constant (5
th
range)
P4.1
Ready indicator / auto
endpoint
endpoint
READY/ON Ready
indicator on; auto endpoint off
P4.2 Select
°C/°F
°C
Degrees
Celsius
P4.3
ATC on or off
ATC on
––
P4.4
TDS factor
0.50
Adjustable from 0.40 to 1.00 (available in TDS
400 and CON 410 meters only)
400 and CON 410 meters only)
P5.1
Temperature coefficient
2.1 % per °
C
C
Adjustable from 0.0 to 10%
P5.2
Normalisation temperature
25° C
Adjustable from 15 to 30° C
P6.1
Cell constant
1.0
Select from k=1.0, 0.1 or 10
P8.0
Factory default
No
Retains your current settings
NOTES:
The P7.0 clock function retains the year and time that you set, even after reverting to factory
default settings.