Non-Invasive Monitoring Systems Inc. ER104054 User Manual

Page of 27
Nims, Inc. 
23 
August 2008 
Exer-Rest Operations Manual 302-026-02 Rev. A 
 
 
Chapter 5 
Trouble Shooting 
Problem 
 
Condition 
 
Action 
WILL NOT WORK 
No Response 
Confirm device is connected to 
mains electrical power 
 
 
No Speed increase when 
Speed button pressed 
Confirm device is turned ON, on/off 
switch should be set to ON position 
 
 
 
Confirm Speed not already at 
maximum setting 
 
 
No Travel increase when 
plus button pressed 
Confirm device is turned ON, on/off 
switch should be set to ON position 
 
 
 
Confirm Travel not already at 
maximum setting 
 
Error Code: 01 
Emergency stop button is 
pressed in 
Release the Emergency stop button 
or rotate it clockwise before 
restarting the device 
 
Error Code: 02 
Driver power is turned off 
because the left limit 
switch is active 
Push the motion frame component of 
the device towards the middle. It will 
not move easily 
 
Error Code: 03 
Drive power is turned off 
because the right limit 
switch is active 
Push the motion frame component of 
the device towards the middle. It will 
not move easily 
 
Error Code: 04 
The driver has no power 
Turn off the main switch for 30 
seconds and re-start it.  If it still does 
not work, please contact customer 
service 
 
Error Code: 11 
Over-current protection 
Turn off the main switch for 3 
minutes and re-start it.  If it still does 
not work, please contact customer 
service  
 
Error Code: 12 
Represents the encoder 
error of the motor 
Turn off the main switch for 3 
minutes and re-start it.  If it still does 
not work, please contact customer 
service 
 
Error Code: 13 
The motor is in over-
speed protection 
Turn off the main switch for 3 
minutes and re-start it.  If it still does 
not work, please contact customer 
service