Insulet Corporation 019 User Manual
151
n
mylife OmniPod System Options and Settings
Time
12-hour or 24-hour clock
Date
MM/DD/YY
DD/MM/YY
MM.DD.YY
DD.MM.YY
DD/MM/YY
MM.DD.YY
DD.MM.YY
YY.MM.DD
YY-MM-DD
YY-MM-DD
Maximum basal rate
30 U/hr
Default is 3.00 U/hr
Default is 3.00 U/hr
Basal programs
7
Basal rate segments
24 per program
Basal rate increment
0.05 units
Temp basal presets
7
Temp basal
%, Units, or Off
30 minutes to 12 hours in
30-minute increments
Default is Off
30 minutes to 12 hours in
30-minute increments
Default is Off
Blood glucose (BG) goal for
BG history
BG history
Upper and lower limits; 70 to 200
mg/dL in 1 mg/dL increments
mg/dL in 1 mg/dL increments
Blood glucose reminder
On or Off; maximum of 4 active at
any one time. Created at time of
bolus entry; reminder can be set
for 30 minutes to 4 hours after
bolus is started, in 30-minute
increments. Default is Off.
any one time. Created at time of
bolus entry; reminder can be set
for 30 minutes to 4 hours after
bolus is started, in 30-minute
increments. Default is Off.
Blood glucose meter sound
On or Off
Default is On
Default is On
Suggested bolus calculator
On or Off
Default is On
Default is On
Target BG value and
correction threshold
correction threshold
8 time segments; 70 to 200 mg/dL
in 1 mg/dL increments
in 1 mg/dL increments
Minimum BG allowed for
use by the suggested bolus
calculator
use by the suggested bolus
calculator
50 to 70 mg/dL in 1 mg/dL incre-
ments
Default is 70 mg/dL
ments
Default is 70 mg/dL
Insulin-to-carb (IC) ratio
8 time segments; 1 to 150 g
carb/U in 1g carb/U increments
carb/U in 1g carb/U increments
Correction factor
8 time segments; 1 to 300 mg/dL
in 1 mg/dL increments
in 1 mg/dL increments
Reverse correction
On or Off
Default is On
Default is On
Duration of insulin action
2 to 6 hours in 30-minute
increments
increments
Bolus increment
0.05, 0.1, 0.5, or 1.0 units
Default is 0.1 units
Default is 0.1 units
Maximum bolus size
30 units
Default is 10 units
Default is 10 units
Extended bolus
%, Units, or Off
30 minutes to 8 hours in
30-minute increments
Default is Off
30 minutes to 8 hours in
30-minute increments
Default is Off