Insulet Corporation 019 Manuale Utente

Pagina di 190
151
mylife OmniPod System Options and Settings 
Time
12-hour or 24-hour clock
Date
MM/DD/YY
DD/MM/YY
MM.DD.YY
DD.MM.YY
YY.MM.DD
YY-MM-DD
Maximum basal rate
30 U/hr
Default is 3.00 U/hr
Basal programs
7
Basal rate segments
24 per program
Basal rate increment
0.05 units
Temp basal presets
7
Temp basal
%, Units, or Off
30 minutes to 12 hours in 
30-minute increments
Default is Off
Blood glucose (BG) goal for 
BG history
Upper and lower limits; 70 to 200 
mg/dL in 1 mg/dL increments 
Blood glucose reminder
On or Off; maximum of 4 active at 
any one time. Created at time of 
bolus entry; reminder can be set 
for 30 minutes to 4 hours after 
bolus is started, in 30-minute 
increments. Default is Off.
Blood glucose meter sound
On or Off
Default is On
Suggested bolus calculator
On or Off
Default is On
Target BG value and 
correction threshold
8 time segments; 70 to 200 mg/dL 
in 1 mg/dL increments 
Minimum BG allowed for 
use by the suggested bolus 
calculator
50 to 70 mg/dL in 1 mg/dL incre-
ments
Default is 70 mg/dL
Insulin-to-carb (IC) ratio
8 time segments; 1 to 150 g 
carb/U in 1g carb/U increments
Correction factor
8 time segments; 1 to 300 mg/dL 
in 1 mg/dL increments
Reverse correction
On or Off
Default is On
Duration of insulin action
2 to 6 hours in 30-minute 
increments
Bolus increment
0.05, 0.1, 0.5, or 1.0 units
Default is 0.1 units
Maximum bolus size
30 units
Default is 10 units
Extended bolus
%, Units, or Off
30 minutes to 8 hours in 
30-minute increments
Default is Off