CVRx BAROSTIMIPG1 User Manual

Page of 45
 
 
 
 
BAROSTIM NEO LEGACY REFERENCE GUIDE 
8-3
 
Precaution:  If any of these 3 situations is observed, a CVRx representative should be contacted 
immediately.  
o
 
Low lead impedance, less than 300 Ohms, may indicate a short in the lead. 
o
 
High lead impedance, greater than 3000 Ohms, may indicate poor lead connection to 
IPG or a lead fracture. 
o
 
Drastic, abrupt or sudden changes in lead impedance may indicate a problem with a 
lead. 
Confirming a Good Telemetry Link 
It is important to confirm that there is a good telemetry link between the Programmer Interface and 
the IPG after selecting the IPG. 
In order to get a good telemetry signal, ensure that the Programmer Interface is placed in its 
upright position with no obstructions between the Programmer Interface and the IPG.  Telemetry 
performance is best when the Programmer Interface front or back side label is facing the IPG and 
is a distance of 2 meters or less from the IPG. 
Check the quality of the telemetry link between the Programmer Interface and the IPG by looking 
at the Link Quality Indicator (Figure 6) on the Software Application screen.  Two or more bars are 
required in order to ensure that parameter updates occur in the IPG when requested. 
      
      
  
Figure 6: Link Quality Indicator (Excellent, Poor, No Link) 
The 9010 Programmer Interface has the option to be hung from an IV pole using the provided 
hook.  This configuration may improve the quality of the telemetry link in situations where the 
link is difficult to maintain. 
Therapy Settings 
Neo Legacy has up to three independently programmable therapies (Therapies 1, 2, 3).  Each therapy has its 
own status window.  Each Therapy status window contains a therapy status indicator, the therapy settings, 
and an “Edit and Test…” button.  Therapy settings can be changed by clicking the “Edit and Test…” 
button. 
Therapies have independent control of the parameters in Table 1. 
Table 1: Neo Legacy Parameter Settings 
Parameter 
Description 
Range of Values 
Pathway 
Determines the position of applied pulses for the therapy. 
Left, Right, Both 
Pulse Width 
Determines the width of the applied pulse. Can be 
configured individually for the left and right pathways. 
15 microseconds to 1205 
microseconds 
Amplitude 
Determines the amplitude of the applied pulses. Can be 
configured individually for the left and right pathways. 
0.8 milliamps to 25.0 
milliamps 
Therapy Frequency 
Determines the frequency of applied pulses except during 
the Rest portion of the Burst Interval. 
10 to 100 pulses per 
second