Rehabtronics Inc. CSST1 User Manual
Device
Use Troubleshooting
General
Troubleshooting
Issue
Recommended
Action(s)
Device
performance
issues
or unanticipated
behavior.
Discontinue
use and contact customer support.
(see
“Contact and Reordering Information” at the end
of
this manual)
Error
Messages
The
following table lists possible error messages that may appear, and the
recommended
actions to take:
DISPLAY
Message Potential Cause(s)
Recommended
Action(s)
Low
Battery on my
StimRay
Battery
is low
Recharge
the Lithium ion battery
High
Temperature
fault
on my
StimRay
internal
heating of
components
Remove
the StimRay from the
electrode
holder, wait until it
cools
down. Try it again. If the
fault
persists contact
Rehabtronics.inc.
Electrode
fault on
my
StimRay
Electrode
not making
good
contact with skin
or
Stimray not making
good
contact with the
electrode
connector.
Remove
the electrode, reapply
the
gel. remove and place back
the
StimRay onto the electrode
holder.
If problem persists change
the
electrodes.
System
Fault on
my
StimRay
Component
failure or
StimRay
not making a
good
contact with the
electrode
holder.
Remove
and place back the
StimRay
unit to the electrode
connector.
If problem persists
replace
the electrode assembly. If
problem
still not resolved, contact
Rehabtronics
inc.
34
DRAFT