displays immediate information about the patient and type of alarm. If an 
alarm continues unattended, a Reminder Page is sent to all pagers in the 
unit. 
 
!
!
!
 
 
 
Pager coverage is limited and designed to cover a care unit. 
Operating range from transmitter to pager is approximately 150 feet 
with no intervening structures (walls, columns, doors, floors, etc.). 
Coverage is not long range. Exact coverage area will be determined 
during initial installation, and you can re-examine the coverage area 
at any time by performing a Transmitter Range Test as described in 
Section 5: Testing and Periodic Maintenance. 
 
{Insert system components diagrams  
(Figure 1-1 Nellcor Manual)} 
 
AlarmView™ Wireless Data Network Components 
 
An AlarmView System includes up to 24 transmitters (one for each 
bedside oximetry monitor), up to 12 pagers, and one programmer. All 
transmitters and pagers within one system operate on the same radio 
frequency and baud rate. Pages will not be received if the pager is out of 
range. The system communicates as indicated in the functional diagram 
below. 
 
{Insert System Block Diagram 
(Figure 10-1 Nellcor Manual)} 
 
AlarmView™ System Functional Diagram  
 
 
List of Components 
 
An AlarmView system consists of the following components. 
 
§ 
Transmitter (up to 24) 
 
§ 
Antenna (1 per transmitter, to 24) 
 
§ 
Interface Cable (1 per transmitter, to 24)