EMBLEM™ S-ICD PROGRAMMER: OPERATIOn
8. Following the optional optimization process, the Acquire Reference S-ECG screen is
displayed (Figure 29). Select the Continue button to acquire a reference S-ECG.
Figure 29: Acquire Reference S-ECG screen
9. Once the Reference S-ECG acquisition process begins, a status screen appears. The
process may take up to one minute, during which the patient should remain still.
During this process, a template of the patient’s baseline QRS complex is stored in the
pulse generator. Select Cancel at any time to end Reference S-ECG acquisition. When
acquisition is complete, select the Continue button.
Programming Therapy Parameters
Once Automatic Setup has been completed, the pulse generator therapy parameters may be
selected.
To set the therapy parameters:
1. Select the Main Menu icon on the Navigation Bar.
2. Select the Implant button.
3. Select the Device Settings icon on the Navigation Bar to display the Device Settings
2. Select the Implant button.
3. Select the Device Settings icon on the Navigation Bar to display the Device Settings
screen (Figure 30).