Fukuda Denshi Co Ltd LX7230KM User Manual

Page of 112
 
17 
6. SpO
2
 Monitoring 
 
When the transmitter is used without the ECG lead cable, it will measure only 
SpO
2
 
WARNING
 
When the SpO
2
 probe (sensor) is in a connector-off condition, the SpO
alarm will not be generated on the receiving monitor. Make sure that the 
SpO
2
 probe (sensor)/ relay cable is securely connected. If the SpO
waveform/numeric data is not displayed, check the patient’s condition and 
pay attention not to miss the connector-off condition. 
 
CAUTION
 
When using the transmitter with only the SpO
2
 sensor cable, ECG and 
respiration measurements on the receiving monitor shall be turned off to 
prevent an erroneous alarm. 
 
There are varieties of SpO
sensors depending on the purpose and intended 
use for LX-7230N and LX-7230KM. Select a proper one and attach it to the 
patient.  
 
CAUTION
 
SpO
sensor for LX-7230N and SpO
sensor for LX-7230KM are not 
compatible. SpO
sensors for LX-7230KM cannot be used on the LX-
7230N transmitter; vice-versa. Make sure to use the correct sensor for 
operation.