 Monitoring (LX-7230N) 
19 
 
CAUTION 
 SpO
sensors for LX-7230N cannot be used on the LX-7230KM. 
 SpO
2
 sensors are not waterproof. Keep away from liquids. 
  Do not pick up the device pulling the sensor or cable part. It may get 
disconnected from the device and the device may be dropped. 
  A message is displayed when the SpO
sensor is disconnected from the 
device. 
  A message is displayed when the device detects that the SpO
sensor is 
disconnected from the patient. Properly attach the SpO
sensor to the 
patient. 
  Do not reuse the single-use SpO
sensor. It may cause incorrect 
measurements. 
 
Read through the instruction of the SpO
2
 sensor as well.
 
 
CAUTION 
The accuracy of SpO
2
 measurement may be influenced by abnormal 
hemoglobin, such as carbon monoxide hemoglobin (COHb) and 
methemoglobin (MetHb). It may be also affected by cardiogreen or 
intravascular dyes. 
 
In addition, the following case may affect the accuracy of SpO
and pulse 
rate measurement. 
  Outside light (direct sunlight, fluorescent, light therapy equipment, 
surgical light, infrared heat ramp, etc.) 
 Hypoperfusion 
  Excessive body motion (patient’s motion) 
  Pigment injected to the blood for testing 
  In case of measurement during receiving CPR treatment 
  Placement of SpO
2
 sensor on limb with a blood-pressure cuff, arterial 
catheter 
  External colorant such as nail polish 
  Abnormally low or high hemoglobin concentration 
 Venous pulse 
 Electrosurgery 
  Influence of electromagnetic waves from other electronics 
  High-intensity radio waves from cell phone