Biolog-id SSTABIOLOGID User Manual

Page of 27
 
Smart Storage Agitator User Manual
 
Version: 1.0 
 
Validation date of the manual: 18/09/2018                                                                                                    
 
   10 of 27
 
The  operating  principle  of  the  RFID  system  is  based  on  a  transponder  (RFID  tags)  and  an 
interrogator  (coupler).  The  latter  is  an  active  radiofrequency  transmitter  device  that  will 
activate  the  RFID  tags  located  in  the  slot  by  providing  them  with  the  energy  they  need  to 
operate. In addition to the energy, the interrogator sends particular commands to which the 
RFID tag responds. A simple command may be to return the matching donation number to a 
unique identifier. 
2.3.4.  Temperature sensor
 
The  temperature  sensor  integrated  in  the  SST-A  is  waterproof.  It  allows  to  measure  the 
temperature of an area. 
 
Fig. Temperature sensor 
Only  the  temperature  of  the  climatic  chamber  (incubator)  is  the  reference.  SST-A  has  no 
claimed performance regarding temperature. 
The  SST-A  gives  the  temperature  only  as  an  indication.  This  function  does  not  provide  the 
safety associated with maintaining the storage temperature of platelet concentrates.
 
2.3.5.  Agitation sensor
 
The Agitation sensor (accelerometer) integrated in the SST-A kit allows to record the shaking 
and stopping phases of the agitator. 
2.4. Hardware and software compatibility 
2.4.1. 
Agitator 
The SST-A is only compatible with the Helmer PF96 agitator. 
For  more  information,  please  contact  the  Biolog-id  Quality  Department  at  the  following 
address: 
2.4.2.  Climate chamber 
The SST-A is compatible with all environmental chambers used for the storage of the PC bags 
if the platelets agitator is not an incubator.