Biolog-id SSTABIOLOGID User Manual
Smart Storage Agitator User Manual
Version: 1.0
Validation date of the manual: 18/09/2018
10 of 27
The operating principle of the RFID system is based on a transponder (RFID tags) and an
interrogator (coupler). The latter is an active radiofrequency transmitter device that will
activate the RFID tags located in the slot by providing them with the energy they need to
operate. In addition to the energy, the interrogator sends particular commands to which the
RFID tag responds. A simple command may be to return the matching donation number to a
unique identifier.
interrogator (coupler). The latter is an active radiofrequency transmitter device that will
activate the RFID tags located in the slot by providing them with the energy they need to
operate. In addition to the energy, the interrogator sends particular commands to which the
RFID tag responds. A simple command may be to return the matching donation number to a
unique identifier.
2.3.4. Temperature sensor
The temperature sensor integrated in the SST-A is waterproof. It allows to measure the
temperature of an area.
temperature of an area.
Fig. Temperature sensor
Only the temperature of the climatic chamber (incubator) is the reference. SST-A has no
claimed performance regarding temperature.
The SST-A gives the temperature only as an indication. This function does not provide the
safety associated with maintaining the storage temperature of platelet concentrates.
claimed performance regarding temperature.
The SST-A gives the temperature only as an indication. This function does not provide the
safety associated with maintaining the storage temperature of platelet concentrates.
2.3.5. Agitation sensor
The Agitation sensor (accelerometer) integrated in the SST-A kit allows to record the shaking
and stopping phases of the agitator.
and stopping phases of the agitator.
2.4. Hardware and software compatibility
2.4.1.
Agitator
The SST-A is only compatible with the Helmer PF96 agitator.
2.4.2. Climate chamber
The SST-A is compatible with all environmental chambers used for the storage of the PC bags
if the platelets agitator is not an incubator.
if the platelets agitator is not an incubator.