4.2. Hazards of RF radiation 
 
The antennas of the electronic system of the SST-A each emit their turn at a 
frequency of 13.56MHz and a power of 200mW (the law framing the design of 
RFID readers prohibits to exceed a power of 2W). 
ELECTROMEDICAL  EQUIPMENT  requires  special  precautions  regarding  EMC. 
The  SST-A  must  be  installed  and  put  into  service  according  to  the  EMC 
information provided by the ACCOMPANYING DOCUMENTS. 
Portable or mobile RF communications devices may affect  ELECTROMEDICAL 
EQUIPMENT 
The use of ACCESSORIES, transducers and cables other than those specified, 
except  for  transducers  and  cables  sold by  the  EQUIPMENT MANUFACTURER 
or EM SYSTEM as replacement parts for internal components, may result in an 
increase in EMISSIONS or a decrease in the IMMUNITY of the DEVICE or EM 
SYSTEM. 
The  DEVICE  or  EM  SYSTEM  should  not  be  used  alongside  other  devices  or 
stacked with them. 
The  DEVICE  or  the  EM  SYSTEMS  may  be  interfered  with  by  other  devices, 
even if they comply with the CISPR EMISSION REQUIREMENTS. 
 
Guidance and manufacturer's declaration - electromagnetic emissions 
The PRD-7150300A is intended for use in the electromagnetic environment specified below. The customer or user of the 
PRD-7150300A should ensure that it is used in such an environment. 
Emissions test 
Conformity 
Electromagnetic environment - guidelines 
RF emission 
CISPR11 
Group 1 
The PRD-7150300A uses RF energy only for its internal 
functions. As a result, its RF emissions are very low and are 
not likely to cause interference in a nearby electronic device. 
RF emission  
CISPR11 
Class B 
The PRD-7150300A is suitable for use in any premises, 
including domestic premises and those directly connected to 
the public low-voltage power supply network supplying 
domestic buildings. 
Harmonic emissions 
IEC 61000-3-2 
Class B 
Voltage fluctuations / 
Flicker 
IEC 61000-3-3 
Suitable