Guidance and manufacturer's declaration - electromagnetic immunity
The PRD-7150300A is intended for use in the electromagnetic environment specified below. The customer or user of the
PRD-7150300A should ensure that it is used in such an environment.
PRD-7150300A should ensure that it is used in such an environment.
immunity test
Test level
IEC 60601
Level of
conformity
electromagnetic environment - guidance
Electrostatic discharge
(DES)
IEC 61000-4-2
(DES)
IEC 61000-4-2
± 6 kV on contact
± 8 kV in the air
± 8 kV in the air
± 6 kV contact
± 8 kV air
± 8 kV air
The floors should be wooden, concrete or
ceramic tiles. If floors are covered with
synthetic materials, the relative humidity
should be at least 30%.
ceramic tiles. If floors are covered with
synthetic materials, the relative humidity
should be at least 30%.
Fast transients in
bursts
IEC 61000-4-4
bursts
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input /
output lines
supply lines
± 1 kV for input /
output lines
± 2 kV for power supply
lines
± 1 kV for input /
output lines
lines
± 1 kV for input /
output lines
The quality of the power supply network
should be that of a typical commercial or
hospital environment.
should be that of a typical commercial or
hospital environment.
Transient overvoltage
IEC 61000-4-5
IEC 61000-4-5
± 1 kV between
phases
± 2 kV between
phase and ground
phases
± 2 kV between
phase and ground
± 1 kV between phases
± 2 kV between phase
and earth
± 2 kV between phase
and earth
The quality of the power supply network
should be that of a typical commercial or
hospital environment.
should be that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
the electrical power
input lines
IEC 61000-4-11
interruptions and
voltage variations on
the electrical power
input lines
IEC 61000-4-11
<5% UT
(> 95% of hollow
UT) for 0.5 cycle
40% UT
(60% dip in UT) for
5 cycles
70% UT
(30% dip in UT) for
25 cycles
<5% UT
(> 95% of hollow
UT) for 5 s
(> 95% of hollow
UT) for 0.5 cycle
40% UT
(60% dip in UT) for
5 cycles
70% UT
(30% dip in UT) for
25 cycles
<5% UT
(> 95% of hollow
UT) for 5 s
<5% UT
(> 95% of hollow UT)
for 0.5 cycle
40% UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(> 95% of hollow UT)
for 5 s
(> 95% of hollow UT)
for 0.5 cycle
40% UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(> 95% of hollow UT)
for 5 s
The quality of the power supply network
should be that of a typical commercial or
hospital environment. If the PRD-7150300A
user requires continued operation during
power mains interruptions, it is recommended
that the PRD-7150300A be powered from an
uninterruptible power supply or a battery.
should be that of a typical commercial or
hospital environment. If the PRD-7150300A
user requires continued operation during
power mains interruptions, it is recommended
that the PRD-7150300A be powered from an
uninterruptible power supply or a battery.
Magnetic field at the
frequency of the
electrical network
(50/60 Hz)
IEC 61000-4-8
frequency of the
electrical network
(50/60 Hz)
IEC 61000-4-8
3 A / m
3 A / m
Magnetic field at the frequency of the
electrical network should have the
characteristic levels of a representative
location in a typical commercial or hospital
environment.
electrical network should have the
characteristic levels of a representative
location in a typical commercial or hospital
environment.
NOTE UT is the AC mains voltage prior to application of the test level.