Biolog-id SSTRBIOLOGID User Manual

Page of 29
 
Smart Storage for fridge : user manual 
Backup date 
12/05/2017 17:14:00 
Version 
0.10 
Author  
Loïc Marouteau 
 
Manual validation date: 19/04/2017
 
 
27 sur 29
 
 
Red drawer space LED 
 
Possible causes 
 
Action 
Data cannot be written  
Move the bag to a different space and try again  
The drawer has been open for longer than 4 
minutes  
Close the drawer  
Communication with processor lost  
Switch to maintenance mode, then restart the 
processor   
 
Red processor LED 
 
Possible causes 
 
 
Action 
CAN bus communication lost  
Switch to maintenance mode and restart the 
processor  
The CAN bus power supply has short-circuited  
Switch to maintenance mode and restart the 
processor 
The processor probe has become disconnected   Connect the temperature probe  
The Ethernet network has become 
disconnected  
Unplug the Ethernet cable and plug it back in again   
Insufficient SD card memory  
Check that a notification has been sent to the third 
party system. Contact the administrator.  
Power supply fault on at least one drawer  
Check that a notification has been sent to the third 
party system. Contact the administrator. 
Battery charger fault  
Check that a notification has been sent to the third 
party system. Contact the administrator. 
When a red LED appears, try to detect the cause of the fault and eliminate it as quickly as 
possible. 
Contact the supplier of your device if it malfunctions. 
Never carry out repairs or modifications without first obtaining authorisation from Biolog-Id. 
Before  carrying  out  any  maintenance  operations,  the  SST-R  must  be  switched  to 
maintenance mode (cf HMI Manual). 
7  Warranty  
Failure to observe any of the recommendations will void the warranty.