Safety  Information 
iRay Technology Co. Ltd.
                                                                        15 
 
 
 
 
 
 
 
 
 
 
 
 
Keep the same load (the same pressure) on detector when 
acquiring image. Or image will be incorrect. 
CAUTION 
 
Do not close to fire, do not use in high temperature 
Do not invert positive and negative pole   
Do not contact with metal in case of short circuit 
Do not insert sharp objects into battery 
Do not beat battery 
Do not stand on battery 
Do not use battery out of rules 
Do not dispose battery or change internal structure 
Do not submerge battery in water, please keep dry in storage 
and do not contact with water in use 
Please charge battery with charger provided by iRay 
Do not mix battery with ones not provided by iRay 
Do not charge battery with broken charger
Charge battery regularly to avoid over discharge failure. 
Do not use battery when it is severe ballooning. 
1.2  Notes for Using 
When  using  device, take  the following  precautions.  Otherwise,  problems  may  occur  and  there 
would be malfunction. 
 
Before exposure 
 
Be sure to check device daily and confirm that it works properly. 
 
B
e sure there is battery installed on Mars1417XF to avoid power off suddenly 
 
Sudden heating of room in cold areas will cause condensation to form on device. 
In this case, wait until condensation evaporates before performing an exposure. 
If it is used while condensation is formed, problems may occur in the quality of 
captured images. When air-conditioner is used, be sure to raise/lower the 
temperature slowly so that difference of temperature in room and device does 
not occur, to prevent condensation. 
 
The detector should be warmed more than 20 minutes before exposure or 
updating gain or defect template.