Safety Information
iRay Technology Co. Ltd.
15
Keep the same load (the same pressure) on detector when
acquiring image. Or image will be incorrect.
acquiring image. Or image will be incorrect.
CAUTION
Do not close to fire, do not use in high temperature
Do not invert positive and negative pole
Do not contact with metal in case of short circuit
Do not insert sharp objects into battery
Do not beat battery
Do not stand on battery
Do not use battery out of rules
Do not dispose battery or change internal structure
Do not submerge battery in water, please keep dry in storage
and do not contact with water in use
and do not contact with water in use
Please charge battery with charger provided by iRay
Do not mix battery with ones not provided by iRay
Do not charge battery with broken charger
.
Charge battery regularly to avoid over discharge failure.
Do not use battery when it is severe ballooning.
1.2 Notes for Using
When using device, take the following precautions. Otherwise, problems may occur and there
would be malfunction.
Before exposure
Be sure to check device daily and confirm that it works properly.
B
e sure there is battery installed on Mars1417XF to avoid power off suddenly
Sudden heating of room in cold areas will cause condensation to form on device.
In this case, wait until condensation evaporates before performing an exposure.
If it is used while condensation is formed, problems may occur in the quality of
captured images. When air-conditioner is used, be sure to raise/lower the
temperature slowly so that difference of temperature in room and device does
not occur, to prevent condensation.
The detector should be warmed more than 20 minutes before exposure or
updating gain or defect template.