Safety  Information 
iRay Technology Co. Ltd.
                                                                        16 
During exposure 
 
Do not move power cable during exposure, or it may cause image noise or 
artifacts, even incorrect images. 
 
Do not use equipment near detector generating strong magnetic field. Otherwise, 
it may cause image noise, artifacts or even incorrect images. 
 
Do not exposure in 60s after 4 times of full range exposure. Otherwise, image 
would be incorrect. Neither do exposure in 30s after a full range exposure. The 
larger dose using, the longer time should be waited before next exposure.   
 
During image acquisition, Mars1417XF should not be influenced by physical or 
electrical way. 
 
If detector receives dose which is close to full range, clipping data lines appear in 
full range area. 
After exposure 
If detector would not be used in 5 days, it is required to take out battery. If battery 
would not be used at long time, it must be charged to 30%~50% every 3 month 
or 50%~70% every 6 month. 
Disinfection and Cleaning 
 
After every examination, wipe patient contact surfaces with disinfectants such as 
ethanol, to prevent the risk of infection. For details on how to sterilize, consult a 
specialist. 
 
Do not spray detector directly with disinfectants or detergents. 
 
Wipe it with cloth slightly damped with neutral detergent. Do not use solvents 
such as alcohol, thinner, benzene, acid and base. Doing so may damage surface 
of detector. 
 
It's recommended to use waterproof non-woven cover as isolated layer between 
detector and blooding patient.