Safety Information
iRay Technology Co. Ltd.
16
During exposure
Do not move power cable during exposure, or it may cause image noise or
artifacts, even incorrect images.
Do not use equipment near detector generating strong magnetic field. Otherwise,
it may cause image noise, artifacts or even incorrect images.
Do not exposure in 60s after 4 times of full range exposure. Otherwise, image
would be incorrect. Neither do exposure in 30s after a full range exposure. The
larger dose using, the longer time should be waited before next exposure.
During image acquisition, Mars1417XF should not be influenced by physical or
electrical way.
If detector receives dose which is close to full range, clipping data lines appear in
full range area.
After exposure
If detector would not be used in 5 days, it is required to take out battery. If battery
would not be used at long time, it must be charged to 30%~50% every 3 month
or 50%~70% every 6 month.
Disinfection and Cleaning
After every examination, wipe patient contact surfaces with disinfectants such as
ethanol, to prevent the risk of infection. For details on how to sterilize, consult a
specialist.
Do not spray detector directly with disinfectants or detergents.
Wipe it with cloth slightly damped with neutral detergent. Do not use solvents
such as alcohol, thinner, benzene, acid and base. Doing so may damage surface
of detector.
It's recommended to use waterproof non-woven cover as isolated layer between
detector and blooding patient.