Olympus Winter & Ibe GmbH ESG200 User Manual
102
Appendix B
Mode
Default
power
setting
Minimum
power
level
1)
Maximum
power
level
1)
Maximum
output
voltage
Rated
load
End-of-
procedure
load
Strong RFITT
+ RCAP
20
1 W
40 W
260 V
p
100 Ω
900 Ω
Mode characteristics for monopolar modes
1)
The power level is the continuous power (Watts) which can be applied during the
activation of the electrosurgical generator. The actual applied power depends on the
tissue characteristics, e.g. resistance.
Characteristics of High Power Cut Support (HPCS)
The following is an example of the characteristics of HPCS during PulseCut mode:
Characteristics of HPCS
P
HPCS
CUT
COAG
t
1
Spark
Spark
time
DRAFT
–
2017-01-17