Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
102
Appendix B
Mode
Default 
power 
setting
Minimum 
power 
level
1)
Maximum 
power 
level
1)
Maximum 
output 
voltage 
Rated 
load
End-of-
procedure 
load
Strong RFITT 
+ RCAP
20
1 W
40 W
260 V
p
100 Ω
900 Ω
Mode characteristics for monopolar modes
1)
 The power level is the continuous power (Watts) which can be applied during the 
activation of the electrosurgical generator. The actual applied power depends on the 
tissue characteristics, e.g. resistance.
Characteristics of High Power Cut Support (HPCS)
The following is an example of the characteristics of HPCS during PulseCut mode:
Characteristics of HPCS
HPCS 
CUT 
COAG 
t
1
Spark 
Spark 
time 
DRAFT 
– 
2017-01-17