Appendix B
APPENDIX B
Mode characteristics according to IEC 60601-2-2
Mode characteristics of monopolar modes
The waveforms of all monopolar modes are continuous, except of ForcedCoag
1)
.
Mode
Default 
power 
setting
Minimum 
power 
level
3)
Maximum 
power 
level
3)
Maximum 
output 
voltage 
Rated 
load
End-of-
procedure 
load
FineCut
30
5 W
120 W
580 V
p
500 Ω
PureCut
30
5 W
120 W
630 V
p
500 Ω
StrongCut
30
5 W
120 W
680 V
p
500 Ω
Pulse Cut
30
5 W
120 W
690 V
p
500 Ω
SoftCoag
30
5 W
120 W
250 V
p
100 Ω
ForcedCoag 
2)
30
5 W
120 W
2000 V
p
500 Ω
Mode characteristics for monopolar modes
1) 
Waveform ........................................................................... Modulated sinus, 15 kHz
2)
 Crest factor at rated load ............................................................................3.0 to 8.0
3)
 The power level is the continuous power (Watts) which can be applied during the 
activation of the electrosurgical generator. The actual applied power depends on the 
tissue characteristics, e.g. resistance.
Mode characteristics of bipolar modes
The waveforms of all bipolar modes are continuous.
Mode
Default 
power 
setting
Minimum 
power 
level
1)
Maximum 
power 
level
1)
Maximum 
output 
voltage 
Rated 
load
End-of-
procedure 
load
FineCut
30
5 W
120 W
470 V
p
500 Ω
PureCut
30
5 W
120 W
530 V
p
500 Ω
StrongCut
30
5 W
120 W
560 V
p
500 Ω
SoftCoag
30
5 W
120 W
260 V
p
100 Ω
Fine RFITT
10
1 W
40 W
250 V
p
100 Ω
dynamic
Pure RFITT
10
1 W
40 W
150 V
p
100 Ω
400 Ω
Strong RFITT 20
1 W
40 W
260 V
p
100 Ω
900 Ω
DRAFT 
– 