Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
35
Installation
4 Installation
4.1 General inspection
Before installing and connecting, inspect the electrosurgical generator and its equipment as 
follows:
 
• Check that the electrosurgical generator, its equipment and all accessories do not have:
 - any dents, cracks, kinks, or deformations
 - any deep scratches
 - any corrosion
 - any missing or loose parts
 
• Check all markings on the electrosurgical generator, its equipment and all accessories for 
clear visibility.
4.2 Placement of the electrosurgical generator
To prevent malfunction or damage of the electrosurgical generator and its accessories take 
account of the following circumstances when placing the electrosurgical generator:
 
• Place the electrosurgical generator on a stable, level surface.
 
• Put the electrosurgical generator upright onto its feet. Do not install the electrosurgical 
generator on its side or upside down.
 
• If the electrosurgical generator is placed on a trolley, the trolley must be of adequate 
strength and size to hold the electrosurgical generator securely.
 
• Make sure that there is sufficient space at the back of the electrosurgical generator to 
allow air circulation at the ventilation slots.
 
• Only use the electrosurgical generator in compliance with the environmental conditions as 
specified in the chapter “Specifications” on page 90.
 
• When lifting up, grip the electrosurgical generator at its sides with both hands. Do not hold 
the electrosurgical generator at its fuse holder.
Interference and automatic instrument recognition
The electrosurgical generator is equipped with a radio transmitting and receiving technology. 
Other radio transmitters and receivers, e. g. mobile phones, may interfere with the automatic 
instrument recognition.
 
• To avoid interference observe the instructions in the chapter “Electromagnetic 
compatibility” on page 92.
4.3 Connecting the electrosurgical generator to the mains electricity
WARNING
Electric shock
In case that a damaged electrosurgical generator is connected inadequately to the mains 
electricity, there is the risk that internal electrical components conduct voltage to the 
housing. This could result in electric shock to the user.
 
• Connect the mains plug of the cord set directly to a grounded fixed socket-outlet.
 
• If a multiple socket-outlet is used take account of the following:
 
• The multiple socket-outlet must be equipped with an insulating transformer of 
protection class I, conforming to IEC 60601-1.
 
• Observe the maximum permitted current or power loading of the multiple socket-outlet 
and the insulating transformer.