Installation
4.4 Connecting the double pedal foot switch
WARNING
Secure connection of the plug
If the foot switch plug is not connected securely, then the energy output may not be 
activated. There is the risk that the HF instrument cuts tissue mechanically. This could result 
in bleeding and perforation of the tissue.
 
• Securely connect the foot switch plug to the foot switch socket at the electrosurgical 
generator as described below.
CAUTION
Not waterproof
The foot switch plug is not waterproof. Any liquid, e.g. water, that gets into the plug can 
damage the foot switch.
 
• Take care that the foot switch plug does not get wet.
CAUTION
Proper disconnection of the plug
Tugging and pulling at the foot switch cable can break the electric circuit and can damage 
the foot switch.
 
• When disconnecting the foot switch plug, directly pull at the plug.
 
• Do not tug or pull at the cable.
CAUTION
Compatible foot switch
If a non-compatible foot switch is connected to the electrosurgical generator, then the foot 
switch might not function correctly. This could cause equipment damage which might result 
in patient injury.
 
• Only use the compatible foot switch (WB50402W) which is within the scope of delivery.
4.4.1 Before connecting
 
• Inspect the foot switch and its cord set for scratches, cracks, and other damages.
Do not use the foot switch if any kind of damage is observed.
 
• Check that the each pedal of the foot switch runs smoothly.
Do not use the foot switch if any kind of problem with the pedals is observed.
4.4.2 Connecting
To connect the double pedal foot switch to the electrosurgical generator proceed as follows:
1.  Correctly align the pins inside the foot switch plug with the receptacles of the foot switch 
socket.