Neutral electrode operation
7.6 Verifying the contact quality monitor indicator
If the indicator shines green although there is no neutral electrode attached, there is a 
malfunction of the electrosurgical generator. In this case, do not use this electrosurgical 
generator.
Figure 7.5  Split type indicator for CQM shining red on the Set Screen and on the 
All Screen
 
• After attaching a neutral electrode to the patient, check the CQM indicator:
When using a split type neutral electrode
The split type indicator must shine green. Refer to the section “Split type and non-split type 
neutral electrodes and CQM” on page 54.
When using a non-split type neutral electrode
The non-split type indicator must shine green. Refer to the section “Split type and non-split 
type neutral electrodes and CQM” on page 54.