 Output settings
1.  On the All Screen tap into the pane for the required output socket, i.e. either 
[MONOPOLAR] or [BIPOLAR].
The Set Screen is displayed.
Figure 8.7  Set Screen for monopolar modes
2.  Tap the yellow button for selecting a cutting mode or tap the blue button for selecting a 
coagulation mode.
The Mode Screen is displayed. The currently selected mode is highlighted.
Figure 8.8  Mode Screen for monopolar cutting modes
3.  Tap the button of the required mode, e.g. [FineCut].
The display changes back to the Set Screen and shows the chosen mode.
4.  Increase or decrease the output power level by tapping [+] or [−] of the chosen mode.
Tapping [Off] on the Mode Screen completely deactivates cutting or coagulation for this 
output socket. Then, instead of a number for the output power level, [--] is displayed.
8.5 The bipolar RFITT coagulation modes
The medical purpose of the radiofrequency induced thermotherapy (RFITT) is to achieve 
controlled tissue coagulation. The electrosurgical generator offers 4 different RFITT modes 
as mentioned in the table above.
One of these modes also offers Resistance Controlled Automatic Power (RCAP) which is 
explained further below.