Inspection before operation
9 Inspection before operation
WARNING
Irregularities and damage
Any irregularity or damage of the electrosurgical generator and its equipment compromises 
the safety and can result in injury to the patient, the user and the medical personnel.
 
• Prior to every operation, inspect the electrosurgical generator as described in this chapter.
 
• Prior to every operation, inspect all equipment for irregularity or damage.
 
• Do not use the electrosurgical generator if any irregularity or damage is suspected.
 
• Do not use the equipment if any irregularity or damage is suspected.
 
• Refer to the chapter “Troubleshooting” on page 78.
9.1 Inspecting the supply of power
Power switch and start screen
To test the power connection of the electrosurgical generator proceed as follows:
1.  Switch on the electrosurgical generator.
2.  Check that the power switch lights up.
3.  Check that the touch screen lights up.
The start screen with the Olympus logo is displayed shortly and the start melody is 
played.
4.  Check that the All Screen is displayed after the start screen.
The All Screen recalls the last settings of all output sockets.
9.2 Inspecting the touch screen
CAUTION
Operation of the touch screen
Wrong operation of the touch screen can cause malfunction or damage to the touch screen.
 
• Only tap one button or pane at a time.
 
• Only tap the touch screen from the front. Avoid oblique angles when tapping the touch 
screen.
 
• Do not use pointed objects, e.g. a pen, to tap the touch screen.
 
• Do not tap the touch screen unintentionally.
 
• Do not apply excessive loads on the surface of the touch screen.
 
• Remove dust and dirt by wiping the touch screen with a cloth.
Testing the setting of output power levels
Figure 9.1  Changing from the All Screen to the Set Screen