Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
72
Operation
 - The following error message is displayed:
Figure 10.1  Error message E0115: "Activation time limit exceeded"
To reactivate the electrosurgical generator proceed as follows:
1.  Release the foot switch or hand switch.
 - If Autostart is used, release the contact between the HF instrument and the tissue.
2.  Then activate again.
 - If Autostart is used, re-establish the contact between the HF instrument and the 
tissue.
The activation time limit does not apply to RFITT modes.
10.2.3 Functional principles of activation
Screens for activation
Activation of power is possible only if the All Screen or the Set Screen is displayed on the 
touch screen.
Activation of cutting and coagulation modes
All monopolar and bipolar cutting modes are activated as follows:
 - By depressing the yellow cut pedal of the foot switch.
 - By pressing the yellow cut trigger directly at the HF instrument.
All monopolar and bipolar coagulation modes are activated as follows:
 - By depressing the blue coagulation pedal of the foot switch.
 - By pressing the blue coagulation trigger directly at the HF instrument.
The output power is activated as long as the foot switch pedal or the hand switch at the HF 
instrument is pressed.