Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
74
Operation
When attempting to activate although the output power level is set to 0, then an alarm tone 
sounds and the following error message is displayed:
Figure 10.4  Error message E0141 "Power set to zero (--)"
 
• Confirm the message by tapping [OK] or wait for a few seconds and then continue.
Activation with Autostart for bipolar SoftCoag
The bipolar SoftCoag mode can be activated via Autostart if this is assigned to the 
corresponding output socket.
 
• Refer to the section “Autostart for the bipolar mode SoftCoag” on page 22 for more 
details on the Autostart feature.
 
• Refer to the section “Foot switch assignment and Autostart” on page 41 for instructions 
on assigning Autostart.
 
• Refer to the section “Autostart Setup” on page 45 for instructions on setting a time delay 
to Autostart.
Stopping activation
The output of power stops as soon the foot switch pedal or the hand switch is released.
When using Autostart for the bipolar mode SoftCoag, the output of power stops as soon as 
the HF instrument is removed from the tissue.
10.3 After use
After each use, immediately perform the procedures as described in the following 
paragraphs. If cleaning is delayed, debris encrustation may become a source of infection. 
Encrustation may also result in malfunction of the electrosurgical generator.
WARNING
Protective clothing for cleaning and disinfection
Tissue debris and reprocessing chemicals represent a risk of infection for the reprocessing 
personnel.
 
• During cleaning and disinfection wear appropriate personal protective clothing such as eye 
wear, face mask, moisture-resistant clothing and surgical gloves that fit properly.
 
• Before leaving the reprocessing area, remove the contaminated protective clothing.
WARNING
Drying
If the electrosurgical generator is stored or used after cleaning while still damp, then this can 
result to electric shock to the patient and the user.
DRAFT 
– 
2017-01-17