Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
77
Reprocessing
3.  In case of heavy soiling, use additional cloths.
4.  Before disinfection, let the surfaces dry or wipe them dry using a lint-free cloth.
Inspection before disinfection
 
• Visually inspect the product thoroughly after cleaning.
The product must be visually clean.
 
• If the are any signs of debris, repeat the cleaning process.
11.3 Disinfection
Suitable disinfectants
Low-level disinfection efficacy has been validated with an alkaline disinfectant based on low 
alcohol (<20%) and quaternary ammonium compounds (Sani-Cloth
®
 Plus manufactured by 
PDI Professional at 3 minutes contact time.)
Material compatibility has been verified with disinfectants based on the following as active 
ingredients:
 - Low alcohol
 - Quaternary ammonium compounds
 - Amine derivatives
 - Sodium hypochlorite
Disinfection procedure
1.  Take a soft, lint-free cloth that is soaked with a suitable disinfectant.
2.  Thoroughly wet the surfaces of the electrosurgical generator.
3.  The surfaces must remain visibly wet for the contact time that is designated by the 
disinfectant's manufacturer.
4.  Use additional cloths to assure the designated continuous wet contact time.
5.  Let the surfaces dry or wipe them dry using a lint-free cloth.
11.4 Other HF equipment
 
• For the reprocessing or disposal of other HF equipment, e.g. neutral electrodes and HF 
instruments, refer to the corresponding instructions for use of these products.
DRAFT 
– 
2017-01-17