Specifications
16 Specifications
Specifications, design and accessories are subject to change without any notice or any 
obligation of the manufacturer.
16.1 Specifications for the CELON ELITE ESG-200
Size
Width ...............................................................................................................295 mm
Height ..............................................................................................................115 mm
Depth ...............................................................................................................400 mm
Volume .......................................................................................................13,570 cm³
Weight ..............................................................................................................6.61 kg
Packaging
Weight of packaging ...........................................................................................1.5 kg
Type of packaging  ..........................cardboard and expanded polypropylene material
Power supply
Voltage range .................................................. AC 100 V to 120 V/AC 220 V to 240 V
Frequency ............................................................................................... 50 Hz/60 Hz
Maximum input power .....................................................................................  650 VA
Fuse .............................................................................................................. T 6.3 AH
Classifications
According to IEC 60601-1, protection against electric shock:
Medical electrical equipment .............................................................................Class I
Applied parts  ................................................................................................. Type CF
Protection against harmful ingress of water .........................................................IPX0
Mode of operation ...............................Continuous operation with intermittent loading
According to Medical Device Directive 93/42/EEC ...................................................IIb
Output
High frequency functions..........................................................Monopolar and bipolar
High frequency ..................................................................................... 356 kHz ±20%
Maximum HF power .......................................................................................... 120 W
All modes ......................25% duty cycle (e.g. activated for 10 s, deactivated for 30 s)
All RFITT modes ............................................................................... 100% duty cycle
RFID technology
The device contains radio-frequency identification (RFID) technology.
Frequency band: ...........................................13,56 MHz (internationally harmonized)
Ambient conditions for operation, storage and transport
Operating temperature ..............................................+10 to +40 °C (+50 to +104 °F)
Storage and transport temperature .............................-25 to +60 °C (-13 to +140 °F)
Relative operating humidity .............................................30 to 85 %, non-condensing
Relative storage and transport humidity ..........................10 to 85 %, non-condensing
Atmospheric operating pressure ........................................................... 70 to 106 hPa
Atmospheric storage and transport pressure ........................................ 50 to 106 hPa
DRAFT 
– 