Olympus Winter & Ibe GmbH ESG200 User Manual
92
Electromagnetic compatibility
17 Electromagnetic compatibility
Le présent appareil est conforme aux CNR d’ISED applicables aux appareils radio exempts
de licence.
L’exploitation est autorisée aux deux conditions suivantes :
1. l’appareil ne doit pas produire de brouillage,
2. l’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est
susceptible d’en compromettre le fonctionnement.
Precautionary measures
Electromedical devices are subject to special precautionary measures concerning
electromagnetic compatibility (EMC). This device is to be used only for the purposes
described in these instructions for use and has to be installed, set up, and operated in
compliance with the EMC guidelines.
Impact of mobile and portable RF communication devices
The emission of radio frequency energy by mobile communication devices may impact the
function of the electromedical device. Operating such mobile communication devices (e.g.
cell phones, GSM phones) in the proximity of the electromedical device is prohibited.
Electrical connections
Electrical connections bearing this warning symbol may not be touched. Connections
between such plugs and sockets may not be established without first implementing ESD
precautionary measures.
ESD precautionary measures
The following are ESD precautionary measures:
• Connect all electrical equipment to be connected to the device to a potential equalisation
system (via PE).
• Use only the equipment and accessories mentioned in these instructions for use.
The staff have to be informed about and trained in ESD precautionary measures.
Guidelines and manufacturer’s declaration – electromagnetic emissions
This device is intended for use in the electromagnetic environment specified below. The customer or the
This device is intended for use in the electromagnetic environment specified below. The customer or the
user of this device should assure that it is used in such an environment.
Emissions test
Emissions test
Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
During standby this device uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
This device is suitable for use in all facilities including those in
residential areas and those directly connected to a public supply
network that also supplies buildings used for residential purposes.
Harmonic
emissions
IEC 61000-3-2
Class B
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
In
compliance