Voltcraft DSO-1062D 2-Channel Oscilloscope, Digital Storage Oscilloscope DSO-1062D User Manual

Product codes
DSO-1062D
Page of 84
Getting Started Guide 
DSO-1062D Digital Oscilloscope User Manual                                             8 
 
3.5  Probe Attenuation Setting 
Probes are of various attenuation factors which affect the vertical scale of the signal. The Probe 
Check function is used to verify  if  the Probe attenuation option matches the attenuation of the 
probe. 
As an alternative method to Probe Check, you can push a vertical menu button (such as the CH 1 
MENU button) and select the Probe option that matches the attenuation factor of your probe. 
Make sure that the Attenuation switch on the probe matches the Probe option in the oscilloscope. 
Switch settings are 1X and 10X.   
When the Attenuation switch is set to 1X, the probe  limits the bandwidth of the oscilloscope to 
6MHz. To use the full bandwidth of the oscilloscope, be sure to set the switch to 10X. 
 
3.6  Self Calibration 
The self calibration routine helps optimize the oscilloscope signal path for maximum measurement 
accuracy. You can run the routine at any time but should always run it if the ambient temperature 
changes by 5℃  or more. For a more accurate calibration, please power on the oscilloscope and 
wait for 20 minutes until it has adequately warmed up. 
To compensate the signal path, disconnect any probes or cables from  the front-panel input 
connectors. Then, push the UTILITY button, select the Do Self Cal option and follow the directions 
on the screen. 
.