:MEASure:SETup:MAX 
Syntax 
:MEASure:SETup:MAX <value> 
:MEASure:SETup:MAX? 
Description 
Set or query the upper limit of the threshold when measuring time, delay and phase 
parameters. The default unit is %. 
 
Parameter 
Name 
Type 
Range 
Default 
<value>  Integer 
7 to 95 
90 
 
Explanation 
  The upper limit, middle point and lower limit of the threshold are used to define the 
time, delay and phase parameters. Setting these values will affect the measurement 
results of the time, delay and phase parameters.
 
  Setting the upper limit lower than the middle point will automatically reduce the 
middle point to keep it lower than the upper limit. 
 
Return 
Format 
The query returns an integer between 7 and 95.   
Example 
:MEASure:SETup:MAX 95 
Related 
Command
 
 
 
:MEASure:SETup:MID 
Syntax 
:MEASure:SETup:MID <value> 
:MEASure:SETup:MID? 
Description 
Set or query the middle point of the threshold when measuring time, delay and phase 
parameters. The default unit is %. 
 
Parameter 
Name 
Type 
Range 
Default 
<value>  Integer 
6 to 94 
50 
 
Explanation 
  The upper limit, middle point and lower limit of the threshold are used to define the 
time, delay and phase parameters. Setting these values will affect the measurement 
results of the time, delay and phase parameters.
 
  The middle point must be lower than the upper limit and larger than the lower limit.
 
Return 
Format 
The query returns an integer between 6 and 94.   
Example 
:MEASure:SETup:MID 89 
Related 