DS1000Z User’s Guide 
9-2 
To Enable Pass/Fail Test 
 
Press Utility  Pass/Fail  Enable to select “ON”. Note that when the time base 
is in XY mode, the Pass/Fail test function can not be enabled.   
 
To start testing, press Enable and select “ON”. Then, press Operate to select “ ” 
to start testing and select “■” to stop testing.   
 
You can select the signal source, set the test mask range, create mask as well as 
save and load the test mask. For details, please refer to the following introductions.   
 
 
To Select Source 
 
Press Source to select the channel (CH1-CH4) to be tested and only channels 
enabled can be selected. During the test, the oscilloscope will judge whether each 
frame of waveform in the source complies with the current test mask and those 
waveforms pass through the mask area (blue area) is considered as failed.   
 
 
Mask Range 
 
Users can define their desired test masks. The test mask is “Screen” by default. 
 
Press Range to enter the mask range setting menu. Press X Mask and Y Mask
rotate 
  and the mask lines appear on the screen as shown in the figure below. 
Press Create Mask to apply the mask currently created. The horizontal and vertical 
adjustment ranges are 0.02 div to 4.0 div and 0.04 div to 5.12 div respectively.