Rigol DS2202A 2-channel oscilloscope, Digital Storage oscilloscope, DS2202A User Manual

Product codes
DS2202A
Page of 284
 
RIGOL 
DS2000A User’s Guide
 
 
7-13  
 
SPI Decoding (Option) 
 
SPI serial bus consists of clock line (SCLK) and data line (SDA). 
 
Host
Device
SDA
SCLK
SDA
SCLK
   
SCLK
SDA
 
 
SCLK: sample the SDA on the clock rising edge or falling edge. 
SDA: denote the data channel. 
   
 
Press Decode1  Decode to select “SPI” and open the SPI decoding function 
menu. 
 
1.  SCLK Setting   
Press SCLK to enter the clock line setting interface. 
  Press Channel to select any channel (CH1 or CH2) as the clock channel.   
  Press Slope to set to sample the SDA on the rising or falling edge of the 
SCLK. 
  Press Threshold to set the threshold of the clock channel. 
 
2.  SDA Setting   
Press SDA to enter the SDA data line setting interface. 
  Press Channel to select any channel (CH1 or CH2) as the data channel. If 
“OFF” is seelcted, this data line will not be set. 
  Press Polarity to set the polarity of the SDA data line to 
  (high) or 
(low). 
  Press Threshold to set the threshold of the data channel.   
 
3.  Data Bits Setting 
Press Data Bits to set the number of bits of each frame of data. The range