Rigol DS2202A 2-channel oscilloscope, Digital Storage oscilloscope, DS2202A User Manual

Product codes
DS2202A
Page of 284
RIGOL 
  DS2000A User’s Guide 
7-14 
available is from 4 to 32.   
 
4.  Endian Setting 
Press Endian to select “LSB” or “MSB” and the default is “MSB”. 
 
5.  Display-related Setting 
Press Format to set the display format of the bus to Hex, Decimal, Binary or 
ASCII. 
Press Offset and use 
  to adjust the vertical display position of the bus. 
Press BusStatus to turn the bus display on or off.   
 
6.  Decoding Table 
The decoding table displays the decoded data, the corresponding line number, 
time and error information on the data line in table format. It can be used to 
observe relatively longer decoded data to solve the problem that some data 
could not be seen clearly on the screen.   
 
Press Event Table  Event Table to select “ON” (note that this operation is 
only available when BusStatus is set to “ON”) to enter the decoding table 
interface as shown in the figure below. If a USB storage device is currently 
connected to the instrument, press Export to export the data table to the 
external USB storage device in CSV format.