Rigol DS1204B 4-channel oscilloscope, Digital Storage oscilloscope, DS1204B User Manual

Product codes
DS1204B
Page of 164
RIGOL 
©  2008 RIGOL Technologies, Inc.
                                                                   
 
  User‟s Guide for DS1000B Series 
5-3 
Delta Time 
Measurement Accuracy   
(Full Bandwidth) 
Single-shot: ± (1 sample interval + 50ppm ×  reading + 
0.6 ns)   
>16 averages: ± (1sample interval + 50ppm ×  reading 
+ 0.4 ns) 
 
Vertical 
A/D converter 
8-bit resolution, each channel samples simultaneously 
Volts/div Range 
2mV/div-10V/div at input BNC 
Offset Range   
± 40V(500mV~10V) 
± 2V(2mV~200mV) 
Analog Bandwidth  60MHz(DS1062/4B) 
100MHz(DS1102/4B) 
200MHz(DS1202/4B) 
Single-shot 
Bandwidth 
60MHz(DS1062/4B) 
100MHz(DS1102/4B) 
200MHz(DS1202/4B) 
Selectable Analog 
Bandwidth Limit 
(typical) 
20MHz 
Lower  Frequency 
Limit (AC -3dB) 
≤5Hz (at input BNC) 
Rise  Time  at  BNC, 
typical 
<1.75ns, <3.5ns, <5.8ns,   
On 200MHz, 100MHz, 60MHz respectively   
DC Gain Accuracy 
2mV/div~5mV/div:  ± 4%  (Sample  or  Average  acquisition 
mode) 
10mV/div~10V/div:  ± 3%  (Sample  or  Average  acquisition 
mode) 
DC  Measurement 
Accuracy,  Average 
Acquisition Mode 
Average of ≥16 Waveforms with vertical position at zero: 
± (DC Gain Accuracy× reading+0.1div+1mV)   
Average of ≥16 Waveforms with vertical position not at zero: 
± [DC  Gain  Accuracy× (reading+  vertical  position)+(1%  of 
vertical position)+0.2div] 
Add 2mV for settings from 1mV/div to 200 mV/div 
Add 50mV for settings from >200mV/div to 10V/div 
Delta Volts 
Measurement 
Accuracy  (Average 
Acquisition Mode) 
Delta  Volts  between  any  two  averages  of  16  waveforms 
acquired under same setup and ambient 
conditions: ± (DC Gain Accuracy× reading + 0.05 div)