Rigol DS1204B 4-channel oscilloscope, Digital Storage oscilloscope, DS1204B User Manual

Product codes
DS1204B
Page of 164
RIGOL 
5-2                                                                                                      ©  2008 RIGOL Technologies, Inc. 
  User‟s Guide for DS1000B Series 
Specifications
 
Acquisition 
Sampling Modes 
Real-Time 
Equivalent-time 
Sampling Rate 
2 GSa/s (half channel 
[1]
1 GSa/s (each channel) 
50 GSa/s
[2]
 
Averages 
N  time  acquisitions,  all  channels  simultaneously,  N  is 
selectable from 2, 4, 8, 16, 32, 64, 128 and 256 
 
Inputs 
Input Coupling 
DC, AC, GND 
Input Impedance 
1MΩ± 2%, in parallel with 18pF± 3pF 
Probe Attenuation Factors 
0.001X, 0.01X, 0.1X, 1X, 2X, 5X, 10X, 20X, 50X, 
100X, 200X, 500X, 1000X 
Maximum Input Voltage 
Maximum Input Voltage of the analog channel: 
CAT I 300Vrms, 1000Vpk; transient overvoltage 
1000Vpk   
CAT II 100Vrms, 1000Vpk   
RP2200 10:1, CAT II 300Vrms   
RP3200 10:1, CAT II 300Vrms   
RP3300 10:1, CAT II 300Vrms 
Time  delay  between  channel 
(typical) 
500ps 
 
Horizontal 
Sample Rate Range 
3.65Sa/s-2GSa/s (Real-Time),   
3.65Sa/s-50GSa/s (Equivalent-time) 
Waveform interpolation 
Sin(x)/x 
Record Length 
16k samples for half channel
[1]
 
8k samples for each channel   
Scan speed Range 
(Sec/div) 
1ns/div~50s/div, DS1202/4B   
2ns/div~50s/div, DS1102/4B 
5ns/div~50s/div, DS1062/4B 
1-2-5 Sequence   
Sample Rate and 
Delay Time Accuracy 
±50ppm (over any≥1ms time interval)