Rigol DS1104Z-S 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z-S User Manual

Product codes
DS1104Z-S
Page of 264
 
RIGOL 
DS1000Z User’s Guide
 
 
5-37  
 
behind the preset address (Write, Read or R/W direction). After this trigger 
condition is selected:   
--press AddrBits to select “7 bits”, “8 bits” or “10 bits”; 
--press Address to set the address value according to the setting in AddrBits 
and the ranges are from 0x0 to 0x7F, 0x0 to 0xFF and 0x0 to 0x3FF respectively;   
--press Direction to select “Read”, “Write” or “R/W” (note: when AddrBits is 
set to “8 bits”, this setting is not available).   
  Data: the trigger searches for the control byte value on the data line (SDA) 
following which there is a reading bit and an acknowledgement bit and then 
searches for the specified data value. When this event occurs, the oscilloscope 
will trigger on the clock edge of the acknowledgement bit behind the data byte. 
After this trigger condition is selected: 
--press Byte Length to set the length of the data and the range is from 1 to 5;   
--press CurrentBit to select the desired data bit and the range is from 0 to 
(Byte Length×8-1); 
--press Data to set the data pattern of the current data bit to X, H or L. 
--press AllBits to set the data pattern of all the data bits to the data pattern 
specified in Data
  A&D: trigger when the “Address” and “Data” conditions are met at the same 
time. After this trigger condition is selected:   
--press AddrBits to select “7 bits” ,“8 bits” or “10 bits”; 
--press Address to set the address value according to the setting in AddrBits 
and the ranges are from 0x0 to 0x7F, 0x0 to 0xFF and 0x0 to 0x3FF respectively; 
--press Byte Length to set the length of the data and the range is from 1 to 5; 
--press CurrentBit to select the desired data bit and the range is from 0 to 
(Byte Length×8-1); 
--press Data to set the data pattern of the current data bit to X, H or L. 
--press AllBits to set the data pattern of all the data bits to the data pattern 
specified in Data
--press Direction to select “Read”, “Write” or “R/W” (note: when AddrBits is 
set to “8 bit”, this setting is not available).   
 
Trigger Mode: 
Press Sweep to select the 
 (page 
) under this trigger type as auto, 
normal or single. The corresponding status light of the current trigger mode turns on.