Rigol DS1104Z-S 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z-S User Manual

Product codes
DS1104Z-S
Page of 264
 
RIGOL 
DS1000Z User’s Guide
 
 
5-9  
 
Edge Trigger 
 
Trigger on the trigger threshold of the specified edge of the input signal.   
 
Trigger Type: 
Press Type to select “Edge”. At this point, the trigger setting information as shown in 
the figure below is displayed at the upper right corner of the screen.   
 
 
Source Selection: 
Press Source to select CH1-CH4 or AC Line as the 
. The current 
trigger source is displayed at the upper right corner of the screen. 
Note: select channel with signal input as trigger source to obtain stable trigger.   
 
Edge Type: 
Press Slope to select the kind of edge of the input signal on which the oscilloscope 
triggers. The current edge type is displayed at the upper right corner of the screen. 
 
: trigger on the rising edge of the input signal when the voltage level meets 
the preset trigger level. 
 
: trigger on the falling edge of the input signal when the voltage level meets 
the preset trigger level. 
 
: trigger on the rising and falling edges of the input signal when the 
voltage level meets the preset trigger level. 
 
Trigger Mode: 
Press Sweep to select the 
 (page 
) under this trigger type as auto, 
normal or single. The corresponding status light of the current trigger mode turns on.   
 
Trigger Setting: 
Press Setting to set the trigger parameters (trigger coupling, trigger holdoff and 
noise rejection) under this trigger type.   
 
Trigger Level:   
Trigger occurs only when the signal reaches the preset trigger level. 
Use TRIGGER    LEVEL to modify the level. An orange trigger level line and the 
trigger mark “ ” appear on the screen and move up and down with the rotation of 
the knob, while at the same time, the trigger level value (such as
at the lower left corner of the screen also changes accordingly. When stopping 
turning the knob, the trigger level line and the trigger mark disappear in about 2 s.