DS1000Z User’s Guide
 
 
6-35  
 
Auto Mode 
 
In this mode, one or more cursors will appear. You can use auto cursor measurement 
to measure any of the 24 waveform parameters. Before using this mode, you need to 
at least enable one auto measurement parameter and the number of cursors will 
change with the measurement parameter enabled.   
 
Press Cursor  Mode  “Auto” and the number of cursors displayed on the screen 
is determined by the measurement parameter enabled (different measurement 
parameter needs different number of cursors). Note that no cursor will be displayed 
if no auto cursor measurement parameter is enabled or the measurement source has 
no input. When the waveform is expanded or compressed horizontally, the cursor will 
move accordingly.   
 
If multiple measurement parameters are turned on later, you can use Auto Item to 
switch among at most five measurement parameters that are turned on last. 
 
The following figure shows the auto measurement of the rise time of sinusoidal 
signal.