Greisinger T-Logg 120W / 0-10 0 - 10 V Temperature Data Logger 603415 Data Sheet

Product codes
603415
Page of 8
E39.0.1X.6C-01                                                 Operating Instruction T-Logg 120                                                                                          page  5 of 8
ALARM LOW
:
The measured value is below the min. alarm limit.
ALARM HIGH
:
The measured value has exceeded the max. alarm limit.
ERROR 1
:
The input signal has exceeded the measuring range of the logger.
ERROR 2
:
The input signal has been fallen below the measuring range of the logger.
ERROR 3
:
The max. display range (9999) of the logger is exceeded.
ERROR 4
:
The min. display range (-1999) of the logger is under-run.
ERROR 7
:
The T-Logg has detected a system fault.
 Remedy:
  Remove battery and wait about 30 minutes. Then insert the battery again.
 If the error message displayed furthermore, please send the logger to the
  manufacturer to repair.
8  Battery service life and recording time
Measuring cycle:
2 sec.
10 sec.
15 min.
5 hours
Recording time:
8.9 hours
44 hours
166 days
Approx. 9 years
Battery service life:
   —   
Approx. 1 year
Approx. 3 years
   —   
Please note!
Short measuring cycles as well as frequently measuring data transfer result in a reduction of
the battery service life!
Even if the T-Logg is connected, power consumption is increased. So it’s important to con-
nect the logger with the USB interface of the PC only as long as necessary!