10 Specification
Input signal:
refer to device type, or in according to type plate
T-Logg 120 .. - 001:
0 – 1 Volt
input resistance = ~ 100 kOhm
T-Logg 120 .. - 002:
0 – 2 Volt
input resistance = ~ 200 kOhm
T-Logg 120 .. - 010:
0 – 10 Volt
input resistance = ~ 500 kOhm
T-Logg 120 .. - 020:
0 – 20 mA
shunt resistance = ~ 70 Ohm
T-Logg 120 .. - 420:
4 – 20 mA
shunt resistance = ~ 70 Ohm
Display range:
-1999 ... 9999 digit
Scaling:
freely programmable via software
Decimal point:
setable to any position
Accuracy: 
± 0.5% FS  
(at nominal temperature)
Display:
LCD display, 10 mm high, 4-digit
Recording interval:
2s to 5h
Measuring value memory: 16,000 values
Memory type:
FILLING MEMORY
:
>> Once the memory is filled with data, the recording will automatically be halted.
RING MEMORY
:
>> The old data will be overwritten in case of memory overflow.
Recording time:
approx. 9 hours up to 9 years, depending on measuring cycle.
Alarm function:
the measured values are monitored at alarm limits.
Alarm limit and alarm delay (0 ... 500 min.) adjustable via interface.
Nominal temperature:
+25 °C
Working temperature:
-25 … +60 °C
Storage temperature:
-30 … +85 °C
Battery:
CR 2032, exchangeable.
Battery service life:
approx. 3 years (if interval is 15 min.), depending on measuring cycle and operating
temperature.
Interface:
serial interface, 3-pin miniature plug
Please note: the input is not isolated from the interface!
Data communication:
via interface converter
Electric connection:
T-Logg 120 K - ...:
approx. 0.5 m silicone connection cable
T-Logg 120 W - ..:
angle-type plug in accordance to DIN43650
Dimensions / housing:
48.5 x 48.5 x 35.5 mm (H x W x D), plug an fixation flap not included.
Housing made of shock resistant plastic, transparent front made of polycarbonate,
splash water-proof IP65.
EMC:
The T-Logg 120 ... have been manufactured in accordance with the regulations con-
cerning EMC (2004/108/EG).
The device meets EN 61326-1:2008
Additional error: <1%
11 Disposal instructions
The device must not be disposed in the regular domestic waste. Send the device directly to us (sufficiently
stamped), if it should be disposed. We will dispose the device appropriate and environmentally sound.