Fluke Fluke 4325162 User Manual

Product codes
4325162
Page of 56
 
Earth/Ground Tester 
 
Setup 
13 
Table 5. Display Descriptions (cont.) 
Function Displays 
Condition 
Note 
 
 
edw049.eps 
Hardware malfunction 
(for example, current 
overload). 
Switch ON/OFF if still 
faulty. 
The symbol may 
appear when using the 
stakeless 
measurement on low 
resistance circuits. 
 
 
edw050.eps 
EE PROM memory 
access malfunction. 
Contact service. 
 
 
edw051.eps 
Internal computation 
malfunction. 
 
 
 
edw052.eps 
Thermal overload. 
Cool thoroughly. 
 flashes on display 
Setup 
Batteries 
Note 
Rechargeable batteries may be used but must be charged outside of the 
instrument. The number of measurements available with these cells will 
typically be different to those available using alkaline cells. 
This Tester is equipped with six 1.5 V batteries IEC LR6 type AA. Replace or recharge 
the batteries if the battery indicator shows 1 or 0 bars.  
To insert the batteries: 
1.  Switch off instrument, see Figure 2. 
2.  Disconnect all test leads. 
3.  Open battery compartment. 
4.  Insert batteries. Always replace the complete set of batteries. 
5.  Close battery compartment.