4.  Start measurement with "START TEST" button. First, "R1" with positive voltage is 
measured on jack "E". After releasing the "START TEST" button "R
2
" is measured 
with negative voltage on jack "E". The respectively higher measured value is 
displayed first. 
5.  The second measured value can be called with "DISPLAY MENU". If the set limit 
value (R LIMIT) is exceeded the limit can also be displayed. 
Compensation of Measuring Lead Resistance 
1.  Call display of R
K
 with button "DISPLAY MENU". 
2.  Short circuit measuring lead as shown in Figure 14. 
3.  Press "START TEST" button. Value R
K
 is stored after the release of the "START 
TEST" button, the display jumps back to voltage measurement. Thereafter, R
K
 is 
subtracted from the actual measured value. Turning the central rotary switch for a 
short moment deletes the line compensation again. 
 
edw026.eps 
Figure 14. Compensation of Measuring Lead Resistance