Fluke Fluke 4325170 Data Sheet

Product codes
4325170
Page of 10
4  Fluke Corporation     1623-2 and 1625-2 GEO Earth Ground Testers
Measuring principle: Current/voltage measurement (with external current clamp)
Measuring voltage
Vm = 48 V ac 
Short-circuit current
> 50 mA
Measuring frequency
128 Hz
Probe resistance (R
S
)
Max. 100 k
Ω
Auxiliary earth electrode resistance (R
H
)
Max. 100 k
Ω
Monitoring of R
S
, and R
H
 with error indicator
Automatic range selection
Measurement is not performed if the current through the current clamp is too low
Stakeless ground loop measurement (
)
Switch position
Resolution
Measuring range
Accuracy
Operating error
R
A
 4-pole    
0.001 
Ω to 0.1 Ω
0.020 
Ω to 199.9 Ω
± (7 % rdg + 3 d)
± (10 % rdg + 5 d)
Measuring principle: Stakeless measurement of resistance in closed loops using two current transformers 
Measuring voltage
Vm = 48 V ac (primary)
Measuring frequency
128 Hz
Noise current (I
EXT
)
Max. I
EXT
 = 10 A (ac) (R
A
 < 20 
Ω)
Max. I
EXT
 = 2 A (ac) (R
A
 > 20 
Ω)
Automatic range selection
The information regarding stakeless ground loop measurements is only valid when used in conjunction with the recommended current clamps at the minimum 
distance specified.