Fluke Fluke 4325170 Data Sheet

Product codes
4325170
Page of 10
ST
E
12MKH
k   Hz
VmA
nFmH
Test
Limit
H
S
ES
E
1625 LCD update
artwork
october 8, 2012
version 2
VA
Reference Image
1:1 Artwork
VA
5  Fluke Corporation     1623-2 and 1625-2 GEO Earth Ground Testers
Temperature ranges
Operating  temperature range
-10 °C to 50 °C (14° F to 122° F)
Storage temperature range
-30 °C to 60 °C (-22° F to 140° F)
Temperature coefficient
± 0.1 % of range/± 0.1 % of reading/°C <18 °C >28 °C 
Type of protection
IP56 for case, IP40 for battery door according to EN60529
Max voltage
!
 socket 
 to socket 
U
rms
 =  0 V
Sockets ” 
 “  to each other in any combination, max.  U
rms
 = 250 V (pertains to misuse)
Safety:
Protection by double and/or reinforced insulation. Max. 50 V to earth per IEC61010-1.  300V CAT II, 
Pollution degree 2
EMC (Emission Immunity)
IEC61326-1: Portable
Quality standard
Developed, designed and manufactured to comply with DIN ISO 9001
External field influence
Complies with DIN 43780 (8/76)
Auxiliary power
6 x 1.5 V alkaline (IEC LR6 or type AA )
Battery life span
With IEC LR6/type AA: typ. 3,000 measurements (R
E
+R
H
 
≤ 1 kΩ)
With IEC LR6/type AA : typ. 6,000 measurements (R
E
 + R
H
 > 10 k
Ω)
Dimensions (WxHxD)
250 mm x 133 mm x 187 mm (9.75 in x 5.25 in x 7.35 in)
Weight
≤ 1.1 kg (2.43 lb) without accessories
7.6 kg (16.8 lb) including accessories and batteries in carrying case
Case material
Polyester
General
Memory
Internal memory storage up to 1500 records accessible via USB port
Measuring function
Interference voltage and frequency, earthing resistance 3- and 4-pole with/without clip-on current transformer, 
resistance 2-pole with ac, 2- and 4-pole with dc
Display
4 digit (2999 Digit) - 7 segment liquid crystal display, with improved visibility
Operation
Central rotary switch and function keys
1625-2 Specifications