Fluke 1503 Insulation measuring device, 500V, 1000 V (+20%, -0%) 2427883 User Manual

Product codes
2427883
Page of 38
 
Insulation Testers 
 
Replacing the Batteries and Fuse   
  
19 
Replacing the Batteries and Fuse 
Replace the fuse and batteries as shown in Figure 11. 
Follow the steps below to replace the batteries. 
XW
Warning 
To avoid shock, injury, or damage to the 
Tester: 
• 
To avoid false readings, which could lead 
to possible electric shock or personal 
injury, replace the batteries as soon as 
the battery indicator (
B
) appears. 
• 
Use ONLY fuses with the amperage, 
interrupt, voltage, and speed ratings 
specified. 
• 
Turn the rotary switch to OFF and remove 
the test leads from the terminals. 
1.
 
Remove the battery door by using a standard 
screwdriver to turn the battery door lock until the 
unlock symbol aligns with the arrow. 
2.
 
Remove and replace the batteries. 
3.
 
Replace the battery door and secure by turning the 
battery door lock until the lock symbol aligns with the 
arrow. 
F315 mA 1000 V
Min interrupt rating
10000 A
bbw15f.eps
Fuse, Fast, 315 mA, 
1000 V, Min Interrupt Rating 
10000 A 
Fluke PN 2279339 
Battery, 1.5 V AA Alkaline, 
NEDA 15A, IEC LR6 
Fluke PN 376756 
Figure 11. Replacing the Fuse and Battery