Fluke 1503 Insulation measuring device, 500V, 1000 V (+20%, -0%) 2427883 User Manual
Product codes
2427883
1507/1503
Users Manual
Users Manual
20
Specifications
Need intro paragraph here
General Specifications
Maximum Voltage Applied to any Terminal ........ 600 V ac rms or dc
Storage Temperature .......................................... -40
°
C to 60
°
C (-40
°
F to 140
°
F)
Operating Temperature....................................... -20
°
C to 55
°
C (-4
°
F to 131
°
F)
Temperature Coefficient...................................... 0.05 x (specified accuracy) per
°
C for temperatures < 18
°
C or > 28
°
C
(< 64
°
F or > 82
°
F)
Relative Humidity ................................................ Noncondensing
0 % to 95 % @ 10
°
C to 30
°
C (50
°
F to 86
°
F)
0 % to 75 % @ 30
°
C to 40
°
C (86
°
F to 104
°
F)
0 % to 40 % @ 40
°
C to 55
°
C (104
°
F to 131
°
F)
Vibration .............................................................. Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
Shock .................................................................. 1 meter drop per IEC 61010-1 2
nd
Edition (1 meter drop test, six sides, oak floor)
Electromagnetic Compatibility............................. In an RF field of 3 V/M, accuracy = specified accuracy (EN 61326-1:1997).
Safety .................................................................. Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and
IEC/EN 61010-1 2
nd
Edition for measurement category IV 600 V (CAT IV)
Certifications ....................................................... CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard IEC/EN 61010-1
2
nd
Edition
Batteries .............................................................. Four AA batteries (NEDA 15A or IEC LR6)
Battery Life .......................................................... Insulation test use: Tester can perform at least 1000 insulation tests with fresh alkaline
batteries at room temperature. These are standard tests of 1000 V into 1 M
Ω
with a duty
cycle of 5 seconds on and 25 seconds off. Resistance Measurements: Tester can perform
at least 2500 earth-bond resistance measurements with fresh alkaline batteries at room
temperature. These are standard tests of 1
at least 2500 earth-bond resistance measurements with fresh alkaline batteries at room
temperature. These are standard tests of 1
Ω
with a duty cycle of 5 seconds on and 25
seconds off.
Size ..................................................................... 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)
Weight ................................................................. 550 g (1.2 lb.)