30 
Specifications 
Temperature ranges 
Operating temperature range: 
0 °C to +35 °C (+32 °F to +95 °F) 
Storage temperature range: 
-20 °C to +60 °C (-4 °F to +140 °F) 
Temperature coefficient: 
±
0.1 % of rdg / °C (below 18 °C and above 28 °C) 
Operating humidity: 
<95 % RH noncondensing 
Operating altitude: 
2000 m 
Climatic class: 
C1 (IEC 654-1), -5 °C to +45 °C, 5 % to 95 % RH 
Protection type 
 
Case: IP56
Battery door: 
IP40 
Electromagnetic compatibility: 
Complies with IEC61326-1: Portable 
Safety: 
Complies with IEC 61010-1: CAT None, Pollution Degree 2 
External voltage: 
U
ext
, max = 24 V (dc, ac < 400 Hz), measurement inhibited 
for higher values 
Noise rejection: 
>120 dB (162/3, 50, 60, 400 Hz) 
Measurement time: 
6 seconds, typical  
Maximum overload: 
250 V
rms
 (pertains to misuse) 
Batteries: 
6 x 1.5 v, AA, LR6 Alkaline 
Battery life span: 
>3000 measurements, typical 
Dimensions: 
240 mm x 180 mm x 110 mm (9.5 in x 7.1 in x 4.4 in) 
Weight with batteries: 
1.49 kg (3.28 lb) 
Memory: 
Internal memory storage up to 1500 records 
accessible via USB port 
RA 3-Pole and 4-Pole ground resistance measurement 
Resolution Measurement 
range  Accuracy 
Operating 
error 
0.001 … 10 
Ω
 0.020 
Ω
 to 19,99 k
Ω
 
±
(2 % rdg + 3 d) 
±
(5 % rdg + 3 d) 
Note 
For 2-pole measurements, connect terminals H and S with the supplied 
connector cable.