58 
Operating Uncertainties per EN 61557 
The Operating Uncertainty shows the maximum possible uncertainty when all influence 
factors E1-E10 are counted. 
 
Volts 
R
Lo
 
EN 61557-
R
ISO
 
EN 61557-
Z
EN 61557-
Δ
EN 61557-
I
Δ
N
 
EN 61557-
R
E
 
EN 61557-
Intrinsic 
Uncertainty A 
0.80 %  1.50 % 
10.00 % 
6.00 % 
1.00 %  5.00 %  3.50 % 
 
Influence 
Quantity 
Volts 
R
Lo
 
EN 61557-
R
ISO
 
EN 61557-
Z
EN 61557-
Δ
EN 61557-
I
Δ
N
 
EN 61557-
R
E
 
EN 61557-
E1 - Position 
0.00 %  0.00 % 
0.00 % 
0.00 %  0.00 %  0.00 %  0.00 % 
E2 - Supply 
Voltage 
0.50 %  3.00 % 
3.00 % 
3.00 %  3.00 %  2.75 %  2.25 % 
E3 - 
Temperature 
0.50 %  3.00 % 
3.00 % 
3.00 %  3.00 %  2.25 %  2.75 % 
E4 - Series 
Interferences 
Voltage 
- -  - - - - 
1.50 
E5 - Resistance 
of the probes 
and auxiliary 
earth electrodes 
- -  - - - - 
4.00 
E6.2 - System 
phase angle 
- -  - 
1.00 
- - - 
E7 - System 
frequency 
0.50 % 
2.50 % 
0.00 % 
E8 - System 
voltage 
2.50 %  2.50 %  2.50 %  0.00 % 
E9 - Harmonics 
2.00 % 
E10 - D.C. 
Quantity 
- -  - 
2.50 
- - -