12 Adjusting of the instrument 
To adjust the instruments according to your needs proceed like follows: 
  Switch instrument off. 
   Press zero button and keep it pressed. Turn the device on. Release the mode button after the 
segment test (             
) finished. The first parameter “SCL“ is displayed. 
 
  If a parameter should be changed press key up or down (                   ) the present setting of the 
parameter will be shown and can be changed via up/down keys.  
The parameter setting is confirmed by pressing 
 . 
  Jump to the next parameter with 
 . 
 
Parameter 
value 
information 
button  
buttons 
 
Offset of sensors   factory setting
: off
 
OFF 
-5.
0 … +5.0 
 Selectable -5.
0 … +5.0 
Scale correction    factory setting
: off = 0%
 
OFF 
-5.
00 … +5.00 
Adjustment in % 
Pressing  
 again stores the settings, the instruments restarts (segment test) 
Please note:   
If there is no key pressed within the menu mode within 2 minutes, the 
adjusting will be cancelled, the entered settings are lost! 
 
13 Notes to the calibration services 
Calibration certificates - DKD-certificates - other certificates: 
If device should be certificated for its accuracy, it is the best solution to return it with the referring sensors to 
the manufacturer. (please state desired test values, e.g. 1bar abs) 
Only the manufacturer is capable to do efficient recalibration if necessary to get results of highest accuracy!
 
14 Error and system messages 
ERR.  1  measuring range has been exceeded 
ERR. 2  measured value has fallen below permitted range 
ERR. 3  display range has been exceeded (>19999) 
ERR. 4  measuring value has fallen below displayable range (< -1999) 
ERR. 7  System fault - the device has detected a system fault (defective or far outside allowable 
ambient temperature range) 
--- 
Sensor error or value could not be calculated 
 
If the symbol 
LOBAT” is displayed at the left side of display, the battery is weak, measuring can be 
continued for a short time. 
If 
“BAT” is displayed in the display the battery is used up and needs to be replaced. Measuring is no 
more possible.